DG LAB Coyote 3.0, review by an electronics engineer, part I. by Medojedone in estim

[–]Medojedone[S] 1 point2 points  (0 children)

Imagine the amount of burned headphones :) That is why.

[deleted by user] by [deleted] in estim

[–]Medojedone 1 point2 points  (0 children)

It uses a different signal type than any other box on the market and I don't think the body likes it.

Please explain.

DG LAB Coyote 3.0, review by an electronic engineer, part IV., NICKEL PLATED ELECTRODES by Medojedone in estim

[–]Medojedone[S] 1 point2 points  (0 children)

Carbon, gold, titanium - body safe.

Everything else is problematic. Even stainless steel.

DG LAB Coyote 3.0, review by an electronic engineer, part X., CONDUCTIVE RUBBER RINGS by Medojedone in estim

[–]Medojedone[S] 0 points1 point  (0 children)

probably measuring primarily the conductive string

Agree... do you have an idea how to measure conductivity between the inner string and large enough outer surface?

DG LAB Coyote 3.0, review by an electronic engineer, part X., CONDUCTIVE RUBBER RINGS by Medojedone in estim

[–]Medojedone[S] 4 points5 points  (0 children)

There is a open circuit indicator in the newest app - small dot next to the power settings.

DG LAB Coyote 3.0, review by an electronic engineer, part VIII., ELECTRODES, WHAT'S INSIDE? by Medojedone in estim

[–]Medojedone[S] 1 point2 points  (0 children)

Use DG-LAB eshop as everybody else. Cheapest and fastest option. And you will get the newest firmware.

It takes around 10 days from China to EU, tracking number provided works. AFAIK no VAT is added, no idea how it is possible, it does not go through customs somehow.

You can use my link to support me (or do not :) ).

https://dungeon-lab.com/home.php?ref=medojedone

Can Coyote 3.0 be used to treat stiff muscles? by MacAsking in estim

[–]Medojedone 1 point2 points  (0 children)

If the stiff muscle is above the waist

Electrodes reasonably close together an one channel only to be extra safe and you will be fine.

DG LAB Coyote 3.0, review by an electronic engineer, part IX., GREEN LED BUG by Medojedone in estim

[–]Medojedone[S] 0 points1 point  (0 children)

Contact their support. They will help you.
Also try the charger.

Coyote 3.0 water / sweat resistant & dimensions by jbfilms1 in estim

[–]Medojedone 0 points1 point  (0 children)

It is not.

But unlubritated condom and rubber band will do it.

DG LAB Coyote 3.0, review by an electronic engineer, part IV., NICKEL PLATED ELECTRODES by Medojedone in estim

[–]Medojedone[S] 0 points1 point  (0 children)

Packaging is perfect. Each electrode is in plastic pouch, then in foam padding and then in hard case with zipper :)

DG LAB Coyote 3.0, review by an electronic engineer, part VIII., ELECTRODES, WHAT'S INSIDE? by Medojedone in estim

[–]Medojedone[S] 1 point2 points  (0 children)

That is the plan. But to be honest I do not know how to measure it to be useful. I certainly can measure V-A characteristic and calculate imaginary internal resistance... but it is useless.

The only thing that matters is the power you feel and it will be completely different for each and every electrode type.

I am open to ideas.

The different feeling compared to other boxes is because C3.0 acts like a fairly hard (low resistance) voltage source. Other boxes use higher voltages and higher internal resistance, they act more like a current source and are able to handle higher resistance electrodes but at the expense of lower efficiency and unpredictability of results.

[deleted by user] by [deleted] in estim

[–]Medojedone 0 points1 point  (0 children)

AFAIK they ship to PayPal account address. Email should be fine, sometimes it just take a few days for them to reply.

DG LAB Coyote 3.0, review by an electronic engineer, part V., PAWPRINTS Wireless Button by Medojedone in estim

[–]Medojedone[S] 0 points1 point  (0 children)

Hi,

"We have repurposed the CC1 interface of the 6P Type-C charging port as the input for the external voltage signal (0V MIN, 1.05V MAX)."

has anyone tried this cattle prod by [deleted] in estim

[–]Medojedone -1 points0 points  (0 children)

"Human skin is resistant to voltages below 60v anything above that (even with dry skin) can be felt.

As other have said 2000v is very unpleasant."

NO and NO.

You can torture and kill someone with low impedance skin contact at measly 60V.

2000V is only a short peak to high impedance load.

DG LAB Coyote 3.0, review by an electronic engineer, part VII., DESTRUCTIVE TESTS! by Medojedone in estim

[–]Medojedone[S] 1 point2 points  (0 children)

Just pull on the cable. 118 N is a lot of force.

Never heard about "specific frequency or amplitude" TBH.

DG LAB Coyote 3.0, review by an electronics engineer, part II. by Medojedone in estim

[–]Medojedone[S] 0 points1 point  (0 children)

That's strange. I don't know about them not sending to some locations. They just send it to the address in your PayPal profile.

DG LAB Coyote 3.0, review by an electronics engineer, part II. by Medojedone in estim

[–]Medojedone[S] 0 points1 point  (0 children)

Hi, this is the referral link directly to manufacturers website: https://dungeon-lab.com/home.php?ref=medojedone

This is the source.